Business Type:Manufacturer
Country/Region:China
Ddu Verified
HOT Rank
6/10
Beijing Read Eagle Technology Company
We are professional supplier of anesthesia machine,ventilator,accessories.
Business Type:Manufacturer
Country/Region:China
Ddu Verified
HOT Rank
6/10
Respiration mode | IPPV, SIMV, SIGN, MANUAL, STANDBY, DEMO |
I:E ratio | 8:1~1:8 |
Tidal Volume | 50ml~1600ml |
Respiratory frequency | 1~65bpm |
PEEP | mechanical |
Waveform | P-t, F-t |
Patient type | Adult,Pediatric |
Display | 10.4 TFT color screen |
Driven mode | Pneumatically Driven Electronically Controlled |
Pressure Setting | 1-60cm H2O |
Trigger Pressure | -20-20CM H2O |
Inspiration Plateau | 0-55%Ti |
Working mode | Semi-Open/Semi-closed/Closed system |
Gas Supply Pressure | O2:0.27Mpa-0.55Mpa N2O:0.27Mpa-0.55Mpa |
1) 10.4" TFT screen (Sharp brand), all parameters are clearly showed.
2) Metal frame, with plastic spray, durable and lasting.
3) Double vaporizer station, selectatec compatible.
4)Classical Flowmeter of 4 tubes, with O2/N2O linkage device.
5) Could be easily upgraded with functions: eg adding Multi-parameter Monitor, CO2 Monitor, AMIS System
6)Use advanced differential pressure sensor (Honeywell brand) to monitor tidal volume, the sensor probe is built inside the absorber, so as to avoid the affect of water vapor on tidal volume and improve the accuracy.
Safety Function
·When airway pressure reach 6kPa, APL valve will automatically release gas.
·When airway pressure reach 6kPa, the Bellows will stop sending gas to patient.
·Alarm parameters (Audible and/or visual signal):
1)Airway Pressure high / low limit
2)FiO2 high/low limit
3)Minute Ventilation Alarm
4)Low O2 Supply Pressure(when O2 Supply Pressure is lower than 0.15Mpa, will arouse Audible Alarm)
5)N2O automatic Cutoff Alarm
·Monitoring parameters (Under Manual and Ventilation):
Tidal Volume / Respiratory Frequency
Minute ventilation / Airway Pressure
Inspiration Plateau / Respiration mode / FiO2
Electrical Specification
·Mains: 100-240V, 50/60Hz
·Battery: 12V DC, minimum 120 min.(if with)
·Power consumption: ≤50W
Optional
·Patient Monitor
·York Device
·CO2 Monitor
·AGSS
·AMIS
·Circle absorber heating function